Rjs Inspector D4000 Auto Optic (FIRMWARE version A.05 Manuel d'utilisateur Page 16

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 81
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 15
Operators Guide Inspector 4000
002-5564 RJS, Minneapolis, MN
9
5.0 Main Menu Selections
5.1 Calibration
Power On
When the unit is powered on, the following Ready Menu will display after a few
seconds:
Verifier Ready
XXX Aperture: YY
XXX = indicates the wavelength (color) of the light represented in nanometers
YY = indicates the aperture size in mils
Example:
660 Aperture: 10
This display indicates that the installed wand uses red
light at 660 Nanometer wavelength and a 10 mil diameter
aperture size.
Press Enter and Select at the same time to bring up the Main Menu:
Scan Calib Setup
Storage Inactive
I4000 Ver.x.xx
x.xx = Indicates the Software Version.
Press Select until the cursor is on CALIB and press Enter.
Inspector 4000
Reflectance XX%
Vue de la page 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 80 81

Commentaires sur ces manuels

Pas de commentaire